Problem | Cause | Solution |
There is leak | T-joint is loose | Tighten t-joint |
Other joints are loose | Tighten joints | |
Sample heating tube is broken | Replace sample heating tube | |
Sample inlet nut is loose | Tighten sample inlet nut | |
Septum is faulty | Replace to new septum | |
Filter located before flow sensor is clogged | Replace to new filter | |
Flow sensor is broken | Replace flow sensor | |
Flow sensor calibration does not work fine | Execute flow sensor calibration | |
Sensitivity drop | Mercury standard solution is faulty | Recreate mercury standard solution |
There is leak | Fix the leak | |
Sample heating tube is faulty | Replace sample heating tube | |
Joints and cell are soiled | Replace or wash Joints and cell | |
Method is not proper | Change method or reduce sample volume | |
Mercury lamp is faulty | Replace to new mercury lamp | |
V6 valve has leak | Replae to new V6 valve | |
PMT is faulty | Replace to new PMT | |
AFS board is faulty | Replace to new AFS board | |
Flow error | Ar and/or Air cylinder gone or valve close | Replace cylinder/open valve |
Clogged by sample heating tube | Replace sample heating tube | |
Mass flow controller is broken | Replace mass flow controller | |
Strange peak shape / no peak | V1 valve has leak | Replace to new V1 valve |
V3 valve has leak | Replace to new V3 valve | |
Maximum
H3 temperature is low (Thermocouple position is not good or SSR/switching power supply is broken) |
Adjust thermocouple position Replace SSR/switching power supply |
|
H3 does not heat up (H3 coil or thermocouple is broken) |
Replace H3 heater coil or thermocouple | |
Mercury collector tube is faulty | Replace mercury collector tube | |
High blank | Sample heating tube is contaminated | Replace to new sample heating tube |
Condensate trap is contaminated | Replace or wash condensate trap | |
Seputum is contaminated | Replace septum | |
Joints and cell are contaminated | Replace or wash Joints and cell | |
Wash solution is contaminated | Replace to new wash solution | |
Syringe is contaminated | Replace to new syringe | |