PE-1000 Troubleshooting Chart

PE-1000 Troubleshooting Chart


Problem  Cause  Solution
There is leak T-joint is loose Tighten t-joint
  Other joints are loose Tighten joints
  Sample heating tube is broken Replace sample heating tube
  Sample inlet nut is loose Tighten sample inlet nut
  Septum is faulty Replace to new septum
  Filter located before flow sensor is clogged Replace to new filter
  Flow sensor is broken Replace flow sensor
  Flow sensor calibration does not work fine Execute flow sensor calibration
Sensitivity drop Mercury standard solution is faulty Recreate mercury standard solution
  There is leak Fix the leak
  Sample heating tube is faulty Replace sample heating tube
  Joints and cell are soiled Replace or wash Joints and cell
  Method is not  proper Change method or reduce sample volume
  Mercury lamp is faulty Replace to new mercury lamp
  V6 valve has leak Replae to new V6 valve
  PMT is faulty Replace to new PMT
  AFS board is faulty Replace to new AFS board
Flow error Ar and/or Air cylinder gone or valve close Replace cylinder/open valve
  Clogged by sample heating tube Replace sample heating tube
  Mass flow controller is broken Replace mass flow controller
Strange peak shape / no peak V1 valve has leak Replace to new V1 valve
  V3 valve has leak Replace to new V3 valve
  Maximum H3 temperature is low
(Thermocouple position is not good or SSR/switching power supply is broken)
Adjust thermocouple position
Replace SSR/switching power supply 
  H3 does not heat up
(H3 coil or thermocouple is broken)
Replace H3 heater coil or thermocouple
  Mercury collector tube is faulty Replace mercury collector tube
High blank Sample heating tube is contaminated Replace to new sample heating tube
  Condensate trap is contaminated Replace or wash condensate trap
  Seputum is contaminated Replace septum
  Joints and cell are contaminated Replace or wash Joints and cell
  Wash solution is contaminated Replace to new wash solution
  Syringe is contaminated Replace to new syringe
     
     
   
 
    • Related Articles

    • MA-3000 Troubleshooting Chart

      Problem  Cause  Solution There is leak Joints are loose Re-tighten all joints; fitting on T-joint and trap may need 1/4 turn with small wrench  Most Common Sources of Leaks Sample heating tube is broken Remove and Install new sample heating tube   ...
    • PE-1000 User's Manual

      Attached is a PDF of the PE-1000 User's instrument and software instruction manual. Maintenance and troubleshooting procedures for the PE-1000 can be found in this manual.
    • MA-3 Solo Troubleshooting Chart

      Problem  Cause  Solution There is leak Teflon joints are loose Tighten joints  *Run leak check from Maintenance Sample heating tube is broken Replace sample heating tube   Inlet O-ring is broken or damaged replace with new O-ring   Flowcell has an ...
    • RA-4 Troubleshooting Chart

      Problem  Cause  Solution There is leak Drain tank is loose Close drain tank   Sample tube has crack Replace sample tube   Reaction tube is loose on measurement cap Fix reaction tube   Reagent tube is loose on measurement cap Fix reagent tube   Fixing ...
    • PE-1000 Lamp Replacement

      The attached PDF may be used to replace, or exchange, the lamp in the PE-1000 instrument. This procedure was taken directly from the manual.  You can find the manual here.