Problem |
Cause |
Solution |
There is leak |
Drain
tank is loose |
Close drain tank |
|
Sample tube has crack |
Replace sample
tube |
|
Reaction
tube is loose on measurement cap |
Fix reaction tube |
|
Reagent tube is loose on
measurement cap |
Fix reagent
tube |
|
Fixing
screws for measurement cap are loose |
Tighten screws |
|
Sampler alignment is not correct |
Adjust OFFSET-ARM |
|
Sample
tube height is wrong |
Replace sample tube or adjust OFFSET-LIFT |
|
Sealing ring is missing, broken,
or dirty |
Wash or Replace with new
sealing ring |
Flow rate drop |
Membrane disc filters are clogged |
Replace membrane filters |
|
Air pump tube is worn out |
Replace air
pump tube |
|
Air
pump rotor is worn out |
Replace air pump rotor |
|
Air pump is broken |
Replace air
pump |
|
Bubbler
is clogged |
Replce bubbler |
Sensitivity
drop |
Standard solution is faulty |
Remake calibration solutions |
|
There
is a leak (verify by running a leak check) |
Find and fix the leak |
|
Reaction tube is soileded |
Replace or wash reaction tube |
|
Sample
tubes are dirty |
Replace or wash sample tubes |
|
SnCl2 is going bad
|
Remake SnCl2
|
|
SnCl2 is not being added to the sample tube (the line
is clogged or reagent pump has a problem)
|
Replace reagent teflon tube, reagent pump tube, reagent pump
rotor or reagent pump |
|
HONH2·HCl is not being added (the line is clogged or reagent pump has a problem)
|
Replace
reagent tube, reagent pump tube, reagent pump rotor or reagent pump |
High blank |
Sample
tube(s) are dirty |
Wash or replace sample tubes |
High Blank Correction |
*Stir bar was dirty
|
Wash or
replace stir bar |
|
H2SO4 has high blank |
Remake or Replace to new trace-metal grade H2SO4 |
|
KMnO4 has high blank |
Remake new
KMnO4
|
|
HONH2·HCl has high blank |
Remake new HONH2·HCl
|
|
Reaction or Reagent tube is contaminated |
Replace or wash reaction and/or reagent tubing |
|
Internal
tube/drain tank is contaminated |
Clean or Replace Internal tube/drain tank |
|
Measurement cap is contaminated |
Clean or Replace
measurement cap |
Strange peak shape / No peak |
There
is a leak |
Fix leak |
|
Reaction tube is dirty
|
Replace
reaction tube |
|
Sample
tube(s) is dirty
|
Replace or wash sample tube |
|
SnCl2 is getting bad |
Recreate SnCl2 |
|
SnCl2 is not added (the line
is clogged or reagent pump has some problem) |
Replace reagent tube, reagent pump tube, reagent pump
rotor or reagent pump |
|
HONH2·HCl is not added (the line is clogged or reagent pump has some
problem) |
Replace
reagent tube, reagent pump tube, reagent pump rotor or reagent pump |
|
V2
line is always open (tube positon is different or V2 valve is broken) |
Change tube positon / replace V2 valve |
* 4500 Model only |
*Forget to put in stirrer or
stirrer motor is broken |
Put in
stirrer/ replace stirrer motor |