Cleaning the Reagent Tubing

Cleaning the Reagent Tubing

If a contaminated reagent is used, or the reagent tubing becomes contaminated, you will need to clean the reagent tubing.

1.  Carry out "Reagent Discharge".



2.  Empty the reagent bottle, rinse with DI water, then add nitric acid (1+10) to the reagent bottle, and execute "Reagent Filling".



3.  Leave the system standing in this state for at least two hours, preferrably longer.

4.  Empty the nitric acid solution in the lines by running the “Reagent Discharge" step (same as in Step 1).

5.  Follow the above steps with DI water in place of the nitric acid to flush out any remaining nitric acid in the line(s).

 
The above procedure is for the teflon reaction tubing.  It is recommended to replace a contaminated reagent pump tube because it is difficult to eliminate contamination from the pump tube due to the type of material it is made of.
 


    • Related Articles

    • RA-4500 Reagent Line Cleaning

      If a contaminated reagent is used, or the reagent tubing becomes contaminated, you will need to clean the reagent tubing.  1.  Carry out "Reagent Discharge".  2.  Empty the reagent bottle, rinse with DI water, then add nitric acid (1+10) to the ...
    • Cleaning the reaction cap and bubbler

      1.  Remove the resin screws fixing the measuring cap, and detach the cap.  2.   Once removed, there are three more screws holding the cap assembly together. Remove those screws and disassemble the cap.   3.  Put the detached parts in a container, ...
    • RA-4500 Cap and Bubbler Cleaning

      1.  Remove the resin screws fixing the measuring cap, and detach the cap.  The digestion side cap does not need to be removed. 2.   Once removed, there are three more screws holding the cap assembly together. Remove those screws and disassemble the ...
    • Cleaning the flowcell on the RA-4300A

      This procedure will explain how to clean the detector flowcell on the RA-43000A system. First, open the optical system box, and remove the four screws holding the cell in place. Remove the tubing from the inlet and exit and remove the flowcell ...
    • RA-4500 Flowcell Cleaning

      This procedure will explain how to clean the detector flowcell on the RA-4500 system. First, open the optical system box, and remove the four screws holding the cell in place. Remove the tubing from the inlet and exit and remove the flowcell assembly ...