RA-4300FG+ Poor Peak Troubleshooting

RA-4300FG+ Poor Peak Troubleshooting

Make these checks in the following order:

Because they are easy to replace, and cheap, the first thing to do is change out the disc filters on the system.

 ① Carry out the leak check. In the case of NG (not good), check the connections of the tubing throughout the system. If leak check passes proceed to next step.

 ② Check the quantities of dripped reagents. In the case of NG (not good), make sure the tubing is not plugged by cleaning the tubing; if the tubing is not plugged replace the reagent pump tube. If the reagents are delivering properly proceed to next step.

 ③ Replace or clean the collector tube.
 
 
 ⑤ Replace the yellow reaction tubing that carries the mercury gas to the detector.  

 ⑥ Remake standard mercury solution and stannous chloride solution, and make a measurement.
 
 ⑦ Clean the sample container and remake the reagents used for the pretreatment.

 ⑧ Replace the Nafion tube. In the event that the testing is still not good after all these steps are taken, inform our service department.

    • Related Articles

    • RA4300FG+ Poor Sensitivity

      The following are the possible causes of poor sensitivity of the calibration curve: ・Poor preparation or deterioration of standard solution ・Error in calculation or input of standard concentration ・System leak        Place an empty tube in position ...
    • RA-5 User's Manual

      The attached PDF is the Model RA-5 User Operational and Maintenance instrument and software manual. The maintenance procedures for the RA-5 can be found in the manual.
    • MA-3000 Troubleshooting Chart

      Problem  Cause  Solution There is leak Joints are loose Re-tighten all joints; fitting on T-joint and trap may need 1/4 turn with small wrench  Most Common Sources of Leaks Sample heating tube is broken Remove and Install new sample heating tube   ...
    • MA-3 Solo Troubleshooting Chart

      Problem  Cause  Solution There is leak Teflon joints are loose Tighten joints  *Run leak check from Maintenance Sample heating tube is broken Replace sample heating tube   Inlet O-ring is broken or damaged replace with new O-ring   Flowcell has an ...