Cleaning Sample Boats

Cleaning Sample Boats

The primary means of cleaning your boats should be by heat treatment.

Carry out the heat treatment of a sample boat according to (1) or (2) described below.

(1) Repeatedly run empty boats using the MA-3 analyzer until the blank value has been stabilized.

(2) Put a boat in a heat-resistant ceramic container such as a crucible, and heat it at 750C for three hours or more using an electric furnace.

Clean boats should be stored in an air tight container to prevent mercury absorption onto the boat surface

If possible, use different boats for high concentrations and for low concentrations.

Occassionally heat treating is not enough.  To clean a dirty sample boat that can not be cleaned by heat treatment use the following steps:

(1) First, soak the boat(s) in a neutral detergent, like 1% Alconox, for immersion cleaning, and carry out ultrasonic cleaning for at least 30 minutes.

(2) Then, remove sticking dirt or other foreign material using a swab or a brush

(3) Next, soak and/or sonicate the boat in a 1% nitric acid solution for 30 minutes

Do not use a stronger Nitric Acid Solution greater than 1%

(4) Rinse it sufficiently with pure water and allow it to dry completely

    • Related Articles

    • Sample Fork Cleaning

      **Please follow the Sample Fork Replacement Procedure to remove the sample fork.** Once the sample fork has been removed you may flush out the sample fork using squirt bottles.  As a general rule, flush it first with a steady stream of IPA for at ...
    • Sample Fork Replacement

      ① Turn the power OFF, open the top cover panel, and open the front cover panel.    ② With the covers open, power on the instrument and start up the MA-3000 software, and open the maintenance screen.   ③ Remove the Teflon joint on the front side of ...
    • MA-3 Solo Sample Fork Replacemenmt

      Remove Sample Fork Assembly from the MA-3 Solo by removing the two screws attached to the sliding rail. Remove the fork portion from the rest of the assembly by removing the two screws in the center of the assembly.    Attach the replacement fork to ...
    • MA-3000 Price List with Pictures

      The attached PDF file contains the part numbers, prices and frequency of replacement for the most common parts for this mercury analyzer. To place an order, please send an email to orders@agssci.com or call us at (979) 320-0052.
    • Cleaning the reaction cap and bubbler

      1.  Remove the resin screws fixing the measuring cap, and detach the cap.  2.   Once removed, there are three more screws holding the cap assembly together. Remove those screws and disassemble the cap.   3.  Put the detached parts in a container, ...