RA7000 with Sanpra 3 Gas Flow Paths
There are 3 different gas flow paths during each sample measurement, below is a picture for each path.
Related Articles
RA-4 Low Flow Error
If you are getting a "Low Flow Error" on your RA-4300A system there are three main likely causes. A leak, a restriction, or a problem with the air pump. It should be noted, if you have recently changed and parts, or have done any maintenance to any ...
RA-5 User's Manual
The attached PDF is the Model RA-5 User Operational and Maintenance instrument and software manual. The maintenance procedures for the RA-5 can be found in the manual.
MA-3 Solo Flow Sensor Replacement
How to install new flow sensor 1. Remove 8 screws and remove front cover 2. Remove old flow sensor 3. Attach new flow sensor. Connect with CN20 and CN24. Put back front cover and check flow rate.
RA-7000A with Sanpra 3 Maintenance Manual
Please see the attached PDF for the Maintenance Manual of the RA7000A with Sanpra 3 autosampler system. Please note the User Operation manual for this system is separate and can be found at this link.
Zeroing the Mass Flow Controller (MFC) on the MA-3000 (ver 2)
This attached PDF will explain the procedure to zero, or re-zero, the mass flow controller, or MFC, of the MA-3000. You will need to disconnect the gas line from the instrument before starting this procedure, but the instrument will remain powered ...