RA-4300FG+ Error Codes

RA-4300FG+ Error Codes



    • Related Articles

    • RA-4 Low Flow Error

      If you are getting a "Low Flow Error" on your RA-4300A system there are three main likely causes.  A leak, a restriction, or a problem with the air pump. It should be noted, if you have recently changed and parts, or have done any maintenance to any ...
    • RA-3 Optic Adjustment

      This procedure should only be done at the direction of the AGS Technical Support Team, and after all other options have been exhausted. ​It is important that the Voltages are adjusted with a very clean cell and end cap glass pieces in place; ...
    • RA-5 User's Manual

      The attached PDF is the Model RA-5 User Operational and Maintenance instrument and software manual. The maintenance procedures for the RA-5 can be found in the manual.
    • RA-4 Series Autosampler turntable belt adjustment

      (1) Please remove the right side cover. You'll see the belt of the turntable. (2) Please loosen 4 screws and 2 nuts (Blue circle). (3) Please adjust belt tension by 2 screws (Red circle). (4)  Re-tighten all the screws and nuts after the belt has ...
    • Model RA-3320 User's Manual

      The attached PDF is the Model RA-3320A and RA-3310A Instrument and software user's manual. Maintenance and troubleshooting procedures for the RA-3320 can be found in this manual.