Related Articles
RA-4 Low Flow Error
If you are getting a "Low Flow Error" on your RA-4300A system there are three main likely causes. A leak, a restriction, or a problem with the air pump. It should be noted, if you have recently changed and parts, or have done any maintenance to any ...
RA-5 User's Manual
The attached PDF is the Model RA-5 User Operational and Maintenance instrument and software manual. The maintenance procedures for the RA-5 can be found in the manual.
RA-3 Optic Adjustment
This procedure should only be done at the direction of the AGS Technical Support Team, and after all other options have been exhausted. It is important that the Voltages are adjusted with a very clean cell and end cap glass pieces in place; ...
RA-4 Series Autosampler turntable belt adjustment
(1) Please remove the right side cover. You'll see the belt of the turntable. (2) Please loosen 4 screws and 2 nuts (Blue circle). (3) Please adjust belt tension by 2 screws (Red circle). (4) Re-tighten all the screws and nuts after the belt has ...
Model RA-3320 User's Manual
The attached PDF is the Model RA-3320A and RA-3310A Instrument and software user's manual. Maintenance and troubleshooting procedures for the RA-3320 can be found in this manual.