RA-4300A Leak

RA-4300A Leak


(1) Run a leak check from the maintenance screen and follow the onscreen instructions.  If it fails the leak check, first ensure that the cap is correctly aligned on the top of the sample tube.  If it is not, follow this procedure before continuing. Assuming the cap is correctly aligned, exit the maintenance screen and proceed to put 5 mL of DI water into a sample tube and place in position 11 and simulate a sample run with the software for position 11.  You will now be watching for bubbles, or lack thereof, throughout the entire run.
      (a)    If bubbling occurs during the "Purge"  step before measurement.  
              --> There is a leak between the V1 valve and the Yellow-marked tube, check the connections at both ends. 
                  Or the V3 valve is not working.  The V3 valve can be checked in the maintenance screen, listen for actuation as
                  you toggle it on and off.

      (b)      If bubbling does not occur in the waste station during the "Rinse" step
                  -->  There is a leak between the discharge side of the air pump and the yellow-marked tube.

      (c)      If bubbling does not occur in the sample tube during measurement check for the following:
                  -->  (i) A crack or chip in the sample tube
                        (ii) foreign material on the sealing gasket of the cap
                        (iii) Leak where the tubing enters the cap, make sure all the tubing is fully inserted into the cap and do not
                              slide easily
                        (iv) Leak at the dehumidifier or dehumidifier drain tank
                        (v) Leak in the flowcell
                             (vi) Faulty V3 Valve

    • Related Articles

    • RA-4 Low Flow Error

      If you are getting a "Low Flow Error" on your RA-4300A system there are three main likely causes.  A leak, a restriction, or a problem with the air pump. It should be noted, if you have recently changed and parts, or have done any maintenance to any ...
    • MA-3 Solo Manual Leak Check

      This manual leak check procedure should be run if the auto leak check fails during self test to help identify the location of the leak If bubbling does not occur, or is weak, in the humidification bottle proceed with the following steps: 3) The flow ...
    • RA-5 User's Manual

      The attached PDF is the Model RA-5 User Operational and Maintenance instrument and software manual. The maintenance procedures for the RA-5 can be found in the manual.
    • MA-3 Solo Flow Sensor Replacement

      How to install new flow sensor   1.     Remove 8 screws and remove front cover 2.     Remove old flow sensor 3.     Attach new flow sensor. Connect with CN20 and CN24. Put back front cover and check flow rate.  
    • MA-3000 (ver 1) Flow Meter Cleaning

      The attached PDF will demonstrate how to disassemble and clean the flow meter on the MA-3000. The orientation when reassembling is important The front of the meter has writing that faces forward, the clear window is oriented towards the front, and ...