Related Articles
PE-1000 Troubleshooting Chart
Problem Cause Solution There is leak T-joint is loose Tighten t-joint Other joints are loose Tighten joints Sample heating tube is broken Replace sample heating tube Sample inlet nut is loose Tighten sample inlet nut Septum is faulty ...
PE-1000 User's Manual
Attached is a PDF of the PE-1000 User's instrument and software instruction manual. Maintenance and troubleshooting procedures for the PE-1000 can be found in this manual.
Cleaning/Replacing Valves on the PE-1000
To replace or clean the valves on the PE-1000, first make certain that the analyzer is powered OFF and has had time to cool down to a safe temperature. You will then open the blue front door and locate the two screws on the far left side of the ...
Cleaning Sample Boats
The primary means of cleaning your boats should be by heat treatment. Carry out the heat treatment of a sample boat according to (1) or (2) described below. (1) Repeatedly run empty boats using the MA-3 analyzer until the blank value has been ...
PE-1000 Lamp Replacement
The attached PDF may be used to replace, or exchange, the lamp in the PE-1000 instrument. This procedure was taken directly from the manual. You can find the manual here.