MA-3 Solo Heater Failure Troubleshooting

MA-3 Solo Heater Failure Troubleshooting




    • Related Articles

    • MA-3 Solo Troubleshooting Chart

      Problem  Cause  Solution There is leak Teflon joints are loose Tighten joints  *Run leak check from Maintenance Sample heating tube is broken Replace sample heating tube   Inlet O-ring is broken or damaged replace with new O-ring   Flowcell has an ...
    • MA-3000 Troubleshooting Chart

      Problem  Cause  Solution There is leak Joints are loose Re-tighten all joints; fitting on T-joint and trap may need 1/4 turn with small wrench  Most Common Sources of Leaks Sample heating tube is broken Remove and Install new sample heating tube   ...
    • MA-3 Solo Sample Fork Replacemenmt

      Remove Sample Fork Assembly from the MA-3 Solo by removing the two screws attached to the sliding rail. Remove the fork portion from the rest of the assembly by removing the two screws in the center of the assembly.    Attach the replacement fork to ...
    • MA-3 Solo Detector Flowcell Removal and Cleaning Instructions

      The attached PDF are the instructions to remove, disassemble, and clean the detector flowcell of the MA-3 Solo instrument.
    • PE-1000 Troubleshooting Chart

      Problem  Cause  Solution There is leak T-joint is loose Tighten t-joint   Other joints are loose Tighten joints   Sample heating tube is broken Replace sample heating tube   Sample inlet nut is loose Tighten sample inlet nut   Septum is faulty ...