MA-3 Solo Combustion Tube Replacement
The attached PDF and video are the instructions for removing and replacing the combustion tube and gold trap on the MA-3 Solo Instrument.
Related Articles
Removing Combustion Tube - MA3000
Removing the Combustion Tube Carry out maintenance of the heated sections only after the power switch has been turned OFF and the heater has cooled down to room temperature or it could result in burns. 1. With the power off remove the top cover and ...
MA-3 Solo Combustion Tube Removal Video
The following link is an instructional video on how to remove the MA-3 Solo Combustion Tube (aka Catalyst Tube). MA-3 Solo Combustion Tube Removal Video
MA-3 Solo Combustion Tube Installation Video
The following link is an instructional video on how to install the MA-3 Solo Combustion Tube (aka Catalyst Tube). MA-3 Solo Combustion Tube Installation Video
MA-3 Solo Sample Fork Replacemenmt
Remove Sample Fork Assembly from the MA-3 Solo by removing the two screws attached to the sliding rail. Remove the fork portion from the rest of the assembly by removing the two screws in the center of the assembly. Attach the replacement fork to ...
MA-3 Solo Troubleshooting Chart
Problem Cause Solution There is leak Teflon joints are loose Tighten joints *Run leak check from Maintenance Sample heating tube is broken Replace sample heating tube Inlet O-ring is broken or damaged replace with new O-ring Flowcell has an ...